Bluelab pH Controller Connect Manual Manuel d'utilisateur Page 9

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 23
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 8
English
9
12.0 Factory settings
The pH controller connect is set to the following values by default:
Note: ‘Pump runtime’ [HH:MM:SS] is the time the pump has dosed for. ‘Product
runtime’ [Hrs] is the duration of time the pH controller connect has been in use for.
Pump runtime and product runtime are never reset, even after pump replacement.
To restore the pH controller connect to the above factory settings at any stage:
Disconnect the power
Press and hold the ‘cal’ button then reconnect the power.
When ‘Restored Factory Defaults’ is displayed on the screen, release the ‘cal
button’.
1
2
• Mode Monitor
• Set pH 5.8
• Alarm On
• HighAlarm 6.5
• LowAlarm 5.6
• Dose Up/Dn Down/Acid
• OnTime 1Sec
• OffTime 10Min
• Backlight 100% On
• Contrast 16
• 2/3pt Cal 2
• EarthLink Off
• Language eng
• pH Calibration none
13.0 Change the display language
Press ‘menu’. Press ‘’ to nd language. Press
‘enter’ to select.
Press ‘’ or ‘’ to nd the required language.
Press ‘enter’ to save the language selection.
Press ‘menu’ to exit the menu.
1
2
3
Vue de la page 8